Carrier gas (c)
0.1-10 l/min
Dilution gas (g)
1-100 ml/min
Accuracy
±0.5 % full scale
DIL 200/400
High Precision Gas Diluter
The ECO PHYSICS dilution system DIL 200/400 is designed to calibrate gas analyzers manually or remotely controlled. The device is equipped with digital mass flow controllers for highest accuracy.
Precise Dilution.
The DIL 200/400 dilution system serves as a straightforward dilution tool. The carrier gas may be zero air from a generator, such as the ECO PHYSICS PAG 003, compressed air or pure nitrogen from a gas bottle or any other suitable gas. The dilution gas can be a highly concentrated calibration gas, a gas mixture or a toxic gas that is rendered in noxious by lowering the concentration by dilution. The DIL 200 is the standard product for one carrier gas and one dilution gas in channel A, while the DIL 400 has a second channel B with a total of four mass flow controllers (MFC). The half 19” DIL units fit perfectly into a 19” rack, alongside with instruments for analysis of greenhouse gases, such as NOx or O₃.
Maintenance.
The DIL 200/400 is designed for continuous unattended operation with no maintenance. For highest precision periodic recalibration of mass flow, controllers are recommended.
| Gas flow regulation: | |
| Carrier gas (c) | 0.1-10 l/min |
| Dilution gas (g) | 1-100 ml/min |
| Accuracy | ±0.5 % full scale |
| Linearity | ±0.3 % full scale |
| Input flow | 10 l/min (max.) |
| Input pressure | 3-9 bar |
| Output flow | 10 l/min (max.) |
| Dilution ratio | 1:10 – 1:1000 |
| Operating specifications | |
| Gas connections | 1/4“ Swagelok (c)
1/8” Swagelok (d) |
| Supply voltage | 100-240 V |
| Power consumption | 100 VA max. |
| Dimensions | ½ 19” rack (4 HU) / 36×21.1×17.3 cm |
| Weight | 8.6 kg |
|
ECO PHYSICS reserves the right to change these specifications without notice. |
|
Weitere Informationen finden Sie in der Produktbroschüre
Sie sind interessiert an unseren Produkten?
Sie benötigen Unterstützung bei der Produktwahl? Kontaktieren Sie uns – wir beraten Sie gerne.
nCLD 6x Series
Kostengünstig durch vereinfachtes Setup
nCLD 8xx Series
Ausgestattet mit der neuesten grafischen Benutzeroberfläche und mehreren Anschlüssen für die Datenübertragung.
nCLD Supreme Series
Anspruchsvoller Analysator mit automatischer Inbetriebnahme und Messung